Table B-3. Verification Test Record for Agilent 6611C
Model Agilent 6611C Report No.____________ Date_____________
Test Description Minimum
Specification
Recorded Results Maximum
Specification
Voltage Programming and Measurement
Low Voltage V
out
5 mV
_______V +5 mV
Front Panel measurement
V
out
2 mV
_____mV V
out
+2 mV
High Voltage V
out
7.991 V _______V 8.009 V
Front Panel measurement
V
out
4.4 mV
_____mV V
out
+4.4 mV
Current Programming and Measurement (High Range)
Low Current I
out
2 mA
_______A 2 mA
High Current I
out
4.9955 A _______A 5.0045 A
Front Panel measurement
I
out
10.5 mA
_____mA I
out
+10.5 mA
Current Measurement (Low Range)
Low Current measurement
I
out
2.5
µ
A _____
µ
AI
out
+2.5
µ
A
High Current measurement
I
out
22.5
µ
A _____
µ
AI
out
+22.5
µ
A
Current Sink Measurement
Low Current Sink measurement
I
sin
k
22.5
µ
A _____
µ
AI
sin
k
+22.5
µ
A
High Current Sink measurement
I
sin
k
11.1 mA
_____mA I
sin
k
+11.1 mA
Table B-4. Verification Test Record for Agilent 66312A or Agilent 6612C
Model Agilent______________________ Report No.____________ Date_____________
Test Description Minimum
Specification
Recorded Results Maximum
Specification
Voltage Programming and Measurement
Low Voltage V
out
10 mV
_______V +10 mV
Front Panel measurement
V
out
3 mV
_____mV V
out
+3 mV
High Voltage V
out
19.980 V _______V 20.020 V
Front Panel measurement
V
out
9 mV
_____mV V
out
+9 mV
Current Programming and Measurement (High Range)
Low Current I
out
1 mA
_______A 1 mA
High Current I
out
1.998 A _______A 2.002 A
Front Panel measurement
I
out
4.5 mA
_____mA I
out
+4.5 mA
Current Measurement (Low Range)
Low Current measurement
I
out
2.5 µA _____µAI
out
+2.5 µA
High Current measurement
I
out
22.5 µA _____µAI
out
+22.5 µA
Current Sink Measurement
Low Current Sink measurement
I
sin
k
22.5 µA _____µAI
sin
k
+22.5 µA
High Current Sink measurement
I
sin
k
4.85 mA
_____mA I
sin
k
+4.85 mA