Agilent Technologies N2781A Guía de usuario Pagina 174

  • Descarga
  • Añadir a mis manuales
  • Imprimir
  • Pagina
    / 278
  • Tabla de contenidos
  • MARCADORES
  • Valorado. / 5. Basado en revisión del cliente
Vista de pagina 173
174 Agilent InfiniiVision 2000 X-Series Oscilloscopes User's Guide
13 Measurements
Base
The Base of a waveform is the mode (most common value) of the lower
part of the waveform, or if the mode is not well defined, the base is the
same as Minimum. The Y cursor shows the value being measured.
Overshoot
Overshoot is distortion that follows a major edge transition expressed as a
percentage of Amplitude. The X cursors show which edge is being
measured (edge closest to the trigger reference point).
Rising edge overshοot =
local Maximum D Top
Amplitude
× 100
Falling edge overshοot =
Base D local Minimum
Amplitude
× 100
Overshoot
local Maximum
local Minimum
Base
Top
Overshoot
Vista de pagina 173
1 2 ... 169 170 171 172 173 174 175 176 177 178 179 ... 277 278

Comentarios a estos manuales

Sin comentarios