24
N6773A
3 dB max THD at 3 dB 6 dB max THD at 6 dB
frequency max frequency frequency max frequency
0.2 Vp-p 125 Hz 1.5% 210 Hz 4.0%
0.4 Vp-p 125 Hz 1.5% 180 Hz 4.0%
1 Vp-p 75 Hz 6.0% 95 Hz 6.0%
2 Vp-p 42 Hz 9.0% 60 Hz 9.0%
20 Vp-p 20 Hz 10.0% 37 Hz 10.0%
N6774A
3 dB max THD at 3 dB 6 dB max THD at 6 dB
frequency max frequency frequency max frequency
0.35 Vp-p 125 Hz 1.0% 200 Hz 1.0%
0.7 Vp-p 125 Hz 1.0% 160 Hz 3.0%
1.75 Vp-p 75 Hz 6.0% 95 Hz 6.0%
3.5 Vp-p 40 Hz 8.5% 55 Hz 8.5%
35 Vp-p 20 Hz 10.0% 37 Hz 10.0%
N6775A 3 dB max THD at 3 dB 6 dB max THD at 6 dB
frequency max frequency frequency max frequency
0.6 Vp-p 120 Hz 1.0% 200 Hz 1.0%
1.2 Vp-p 120 Hz 1.0% 160 Hz 3.0%
3 Vp-p 70 Hz 5.0% 95 Hz 6.0%
6 Vp-p 40 Hz 8.5% 55 Hz 8.5%
60 Vp-p 20 Hz 10.0% 35 Hz 10.0%
N6776A
3 dB max THD at 3 dB 6 dB max THD at 6 dB
frequency max frequency frequency max frequency
1 Vp-p 75 Hz 1.0% 160 Hz 1.0%
2 Vp-p 75 Hz 1.0% 150 Hz 3.0%
5 Vp-p 55 Hz 4.0% 75 Hz 6.0%
10 Vp-p 35 Hz 8.0% 45 Hz 8.0%
100 Vp-p N/A N/A 35 Hz 8.0%
N6770 300 W basic DC power modules
Agilent N6705 DC Power Analyzer Supplemental Characteristics (Continued)
Supplemental characteristics are not warranted but are descriptions of performance determined either by design or by type testing.
All supplemental characteristics are typical unless otherwise noted.
Arbitrary waveform generator maximum bandwidth
Maximum bandwidth is based on a sine wave into a resistive load and applies to any output current level. In the tables below, THD means
Total Harmonic Distortion, Vp-p means volts peak-to-peak, 3 dB max frequency is the frequency where the output voltage drops to 3 dB below
the programmed value, and 6 dB max frequency is the frequency where the output voltage drops to 6 dB below the programmed value.