Typical technical information
Gate pulse width: 10 ns to 1 µs
Gate pulse voltage : –4.5 V to 4.5 V
Drain pulse
measurement current: Max. 80 mA
Drain pulse measurement
current resolution: 1 µA
Drain voltage range: –10 V to 10 V
Block diagram for
ultra-short pulsed IV
This block diagram shows a basic
overview of the pulsed IV measure-
ment. The drain voltage (or calculated
current going into the drain) of the
MOSFET is measured by the oscillo-
scope while an ultra-short pulse
is applied to the gate by the PGU.
The benefi ts of this for each
technology type are:
1. For SOI transistors, the short
duty cycle reduces self-heating
effects that adversely impact the
measurement results.
2. For high-k gate dielectric transis-
tors, the short duty cycle reduces
the incidence of electron trapping
that distorts measurement results.
Scope
Bias-TSMU
PGU
50Ω
System confi guration
Operation environment
Supported instrument
Agilent B1500A Semiconductor Device Analyzer
Agilent 4156B/4156C Precision Semiconductor Parameter Analyzer
Agilent 4155B/4155C Semiconductor Parameter Analyzer
Agilent E5270B 8 slot Precision Measurement Mainframe
Agilent E5260A 8 slot High Speed Measurement Mainframe
Agilent 8110A/81101A/81110A Pulse Pattern Generator
Agilent DSO90604A/90404A/90254A/ Infi niium Oscilloscope
DSO80000B series/54850 series/
DSO8000A series/MSO8000A series
Product/option Description
B1542A Pulsed IV Package for B1500/EasyEXPERT
B1542A-001 Pulsed IV SW LTU Media&Manual
B1542A-003 Pulsed IV Option for 4155/4156/E5270/E5260
B1542A-021 Pulsed IV Gate Cable Set
B1542A-022 Pulsed IV Drain Cable Set
B1542A-023 Pulsed IV Docking I/F
B1542A-024 Pulsed IV Switch Set
B1542A-025 Rackmount Kit for Pulse/DC Switch Units
E3333A-002 Add 81110A with 1 output channel
E3333A-015 Add DSO90254A 2.5 GHz Infi niium scope
PGU: 81110A
Scope: DSO90254A
SMU: B1500A