Agilent Technologies E1465A Manual de servicio

Busca en linea o descarga Manual de servicio para Multímetros Agilent Technologies E1465A. Agilent Technologies E1465A Service manual Manual de usuario

  • Descarga
  • Añadir a mis manuales
  • Imprimir
  • Pagina
    / 72
  • Tabla de contenidos
  • MARCADORES
  • Valorado. / 5. Basado en revisión del cliente
Vista de pagina 0
Agilent 75000 SERIES C
Agilent
E1465A/E1466A/E1467A
Matrix Switches
Service Manual
The information in this manual applies directly to:
E1465A 16x16 Matrix Switches
E1466A 4x64 Matrix Switches
E1467A 8x32 Relay Matrix Switches
Copyright© Agilent Technologies, Inc., 1996 - 2006
*E1465-90011*
E1465-90011
Printed: September 2012
E
dition 2
Printed in Malaysia E0912
Vista de pagina 0
1 2 3 4 5 6 ... 71 72

Indice de contenidos

Pagina 1 - *E1465-90011*

Agilent 75000 SERIES CAgilentE1465A/E1466A/E1467A Matrix SwitchesService Manual The information in this manual applies directly to:

Pagina 2

Notes8 Agilent E1465A, E1466A, E1467A Matrix Switch Modules Service Manual

Pagina 3 - Contents

Chapter 1General InformationIntroduction This manual contains information required to test, troubleshoot, and repairthe Agilent E1465A 16 x 16 Relay M

Pagina 4

Relay Life Electromechanical relays are subject to normal wear-out. Relay life dependson several factors. The effects of loading and switching frequen

Pagina 5

NOTERelays that wear out normally or fail due to misuse should not beconsidered defective and are not covered by the product’s warranty.SafetyConsider

Pagina 6 - Safety Symbols

WARNINGREMOVE POWER IF POSSIBLE. Some procedures in thismanual may be performed with power supplied to themainframe while protective covers are remove

Pagina 7 - Declaration of Conformity

Inspection/ShippingThis section contains initial (incoming) inspection and shipping guidelinesfor the matrix modules.InitialInspectionUse the steps in

Pagina 8

Notify Agilent and carrier.Return Module to AgilentNotify AgilentFigure 1-2. Initial (Incoming) Inspection Guidelines14 General Information Agilent

Pagina 9

ShippingGuidelinesFollow the steps in Figure 1-3 to return a matrix module to a AgilentTechnologies Sales and Support Office or Service Center. 1 Pre

Pagina 10

Environment The recommended operating environment for the matrixes is:Environment Temperature HumidityOperating 0oC to +55oC <65% relative (0oC to

Pagina 11 - General Information

Matrix OptionsThe Agilent E1467A Option 211 provides a special matrix expansionterminal module. No other Options to the matrixes are available.Schemat

Pagina 13 - Considerations

18 General Information Agilent E1465A/66A/67A Service Manual

Pagina 14

Chapter 2Verification TestsIntroduction The three levels of test procedures described in this chapter are used toverify that the Agilent E1465A/E1466A

Pagina 15 - Inspection/

Matrix ModuleFunctionalVerificationThe Functional Verification Test for the Agilent E1465A/E1466A/E1467Amatrix modules consists of sending the *TST? c

Pagina 16 - Notify Agilent

The performance verification tests have two parts: a closed channelresistance test of all relay contacts (Test 2-1) and a DC isolation test (Test2-2).

Pagina 17 - Guidelines

Figure 2-2. Agilent E1466A Test FixtureFigure 2-3. Agilent E1467A Test Fixture22 Verification Tests Agilent E1465A/66A/67A Service Manual

Pagina 18 - Description

Test 2-1: Closed ChannelResistance TestThis test verifies that all relay contacts meet the closed-channel resistancespecification for the module. When

Pagina 19 - Test Equipment

2. Ensure all relays are open• Send *RST to Module• Trigger the DMM with TRIG SGL and note reading• DMM reading should indicate an open condition (&g

Pagina 20

Example: ClosedChannel ResistanceTestThis example performs a Closed Channel Resistance Test to measure thematrix relay contact resistances. If the rel

Pagina 21 - Verification Tests

130 PRINT " 1. Turn Mainframe and Agilent 3458A DMM power OFF"140 PRINT " 2. Connect GPIB cable between mainframe and DMM"150 PR

Pagina 22

540 IF Opc <> 1 THEN 520550 ! Check for stuck relays560 OUTPUT @Dmm;"TRIG SGL"570 ENTER @Dmm;A580 IF A<1.E+6 THEN590 PRINT "

Pagina 23 - Test Fixture

ContentsAgilent E1465A/E1466A/E1467A Service ManualChapter 1. General Information . . . . . . . . . . . . . . . . . . . . . .

Pagina 24

Typical ResultHI Path LO PathRow 0 to Col 0 .4162 Ohms .4047 OhmsRow 0 to Col 1 .4045 Ohms .4151 OhmsRow 0 to Col 2 .3947 Ohms .3875 Ohms•ù••ù••

Pagina 25 - Resistance Test

Test 2-2: DCIsolation TestThis test verifies that sufficient DC isolation exists at various points on theModule. DC Isolation is checked from HI to LO

Pagina 26

5. Record the DMM reading in Table 2-1 (HI to LO)HI to Chassis Isolation 1. Make hardware connections as shown in Figure 2-72. Set DMM to 2-wire oh

Pagina 27 - Channel Resistance

5. Record the DMM reading in Table 2-1 (LO to Chassis)Figure 2-8. LO to Chassis Isolation Test Connections10! RE-SAVE "DC_ISOL" 20 ASSIGN

Pagina 28

180 PAUSE 190 OUTPUT @Sw;"*CLS"200 OUTPUT @Sw;"SYST:CDES? "&Cc$210 ENTER @Sw;A$220 SELECT A$[2,2]230 CASE "1"2

Pagina 29

Example:DC Isolation TestThis example performs DC Isolation Tests for HI to LO, HI to Chassis, andLO to chassisTypical ResultDC Isolation TestsHI to L

Pagina 30 - Typical Result

MeasurementUncertaintyFor the performance verification tests in this manual, measurementuncertainties are calculated based on the Agilent 3458A Digita

Pagina 31 - Isolation Test

Table 2-1. Performance Test Record (Page 1 of 10)General InformationTest Facility:Name_____________________________________Address____________________

Pagina 32 - Ω range

Table 2-1. Performance Test Record (Page 2 of 10)Test No/DescriptionMinimum* ValueMeasured Value MaximumValueMeasUncert Test AccRatio (TAR)Test 2-1: C

Pagina 33

Table 2-1. Performance Test Record (Page 3 of 10)Test No/DescriptionMinimum* ValueMeasured Value MaximumValueMeasUncert Test AccRatio (TAR)Test 2-1: C

Pagina 34

Chapter 4. Service . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 55I

Pagina 35 - Test Record

Table 2-1. Performance Test Record (Page 4 of 10)Test No/DescriptionMinimum* ValueMeasured Value MaximumValueMeasUncert Test AccRatio (TAR)Test 2-1: C

Pagina 36 - Ratio (TAR)

Table 2-1. Performance Test Record (Page 5 of 10)Test No/DescriptionMinimum* ValueMeasured Value MaximumValueMeasUncert Test AccRatio (TAR)Test 2-1: C

Pagina 37 - Test Equipment Record

Table 2-1. Performance Test Record (Page 6 of 10)Test No/DescriptionMinimum* ValueMeasured Value MaximumValueMeasUncert Test AccRatio (TAR)Test 2-1: C

Pagina 38 - Test No/Description

Table 2-1. Performance Test Record (Page 7 of 10)Test No/DescriptionMinimum* ValueMeasured Value MaximumValueMeasUncert Test AccRatio (TAR)Test 2-1: C

Pagina 39

Table 2-1. Performance Test Record (Page 8 of 10)Test No/DescriptionMinimum* ValueMeasured Value MaximumValueMeasUncert Test AccRatio (TAR)Test 2-1: C

Pagina 40

Table 2-1. Performance Test Record (Page 9 of 10)Test No/DescriptionMinimum* ValueMeasured Value MaximumValueMeasUncert Test AccRatio (TAR)Test 2-1: C

Pagina 41

Table 2-1. Performance Test Record (Page 10 of 10)Test No/Description MinimumValueMeasured Value Maximum*ValueMeasUncert Test AccRatio (TAR)Test 2-2 D

Pagina 42

Chapter 3Replaceable PartsIntroduction This chapter contains information for ordering replaceable parts for theAgilent E1465A/E1466A/E1467A Relay Matr

Pagina 43

ReferenceDesignatorAgilent PartNumberQty Part DescriptionA2A1 E1466-66501 1 PC ASSEMBLY DENSE MATRIX C101-C106 0160-4835 15 CAPACITOR-FXD 0.1uF +

Pagina 44

ReferenceDesignatorAgilent PartNumberQty Part DescriptionK800-K815 0490-1782 RELAY 2C 12VDC-COIL 1A 250VDC K900-K915 0490-1782 RELAY 2C 12VDC-COIL

Pagina 45

CertificationAgilent Technologies certifies that this product met its published specifications at the time of shipment from the factory. AgilentTechn

Pagina 46

ReferenceDesignatorAgilent PartNumberQty Part DescriptionU111 1820-6475 3 IC FIFO CMOS ASYNCHRO 4608-BIT U112 E1466-88801 1 IC-PAL16L8-25 UNPROG

Pagina 47 - Replaceable Parts

Figure 3-1. Component Assembly Replaceable PartsAgilent E1465A/66A/67A Service Manual Replaceable Parts 49

Pagina 48

Figure 3-2. Terminal Case Assembly Replaceable PartsTable 3-2. Terminal Case Replaceable Parts ReferenceDesignatorAgilent PartNumberQty. Part Descri

Pagina 49

Figure 3-3. Terminal Case Assembly Replaceable PartsTable 3-3. Terminal Case Replaceable PartsReferenceDesignatorAgilent PartNumberQty Part Descript

Pagina 50

Table 3-4. Agilent E1465A Terminal Module Replaceable PartsReferenceDesignatorAgilent PartNumberQty Part DescriptionA3 E1465-66510 1 TERMINAL CARD 1

Pagina 51

Figure 3-4. Agilent E1465A Terminal ModuleFigure 3-5. Agilent E1466A Terminal ModuleFigure 3-6. Agilent E1467A Terminal ModuleAgilent E1465A/66A/67

Pagina 52

Table 3-6. Matrix Module Reference DesignatorsReference DesignatorsA...assembly PCB...printed circuit boardBRK...bracket

Pagina 53

Chapter 4ServiceIntroduction This chapter contains service information for the Agilent E1465A/E1466A/E1467A matrix modules, including troubleshooting

Pagina 54

Troubleshooting To troubleshoot a matrix module problem you must first identify theproblem and then isolate the cause of the problem to a replaceable

Pagina 55

Testing theAssemblyYou can use the tests and checks in Table 4-2 to isolate the problem. SeeFigures 3-1, 3-2, 3-3, 3-4, 3-5, and 3-6 in Chapter 3 for

Pagina 56 - Reference Designators

Frame or chassis ground terminal—typicallyconnects to the equipment’s metal frame.Alternating current (AC).Direct current (DC). Indicates hazardous vo

Pagina 57 - Chapter 4

Relay DriverDescriptionFigure 4-1 shows a simplified diagram of the relay driver and decodercircuits. As shown in the figure, the relays on the compon

Pagina 58 - Identifying

Matching Relaysto ChannelsUse Table 4-3 to help match channel numbers to relay and drive circuitreference designators.Table 4-3. Channel Relays/Refere

Pagina 59 - Assembly

Agilent E1465A Agilent E1466A Agilent E1467ACH # Relay RelayDriverBankDecoderCH # Relay RelayDriverBankDecoderCH # Relay RelayDriverBankDecoderR02 C00

Pagina 60 - Relay Driver

Agilent E1465A Agilent E1466A Agilent E1467ACH # Relay RelayDriverBankDecoderCH # Relay RelayDriverBankDecoderCH # Relay RelayDriverBankDecoderR04 C00

Pagina 61 - Matching Relays

Agilent E1465A Agilent E1466A Agilent E1467ACH # Relay RelayDriverBankDecoderCH # Relay RelayDriverBankDecoderCH # Relay RelayDriverBankDecoderR06 C00

Pagina 62

Agilent E1465A Agilent E1466A Agilent E1467ACH # Relay RelayDriverBankDecoderCH # Relay RelayDriverBankDecoderCH # Relay RelayDriverBankDecoderR08 C00

Pagina 63

Agilent E1465A Agilent E1466A Agilent E1467ACH # Relay RelayDriverBankDecoderCH # Relay RelayDriverBankDecoderCH # Relay RelayDriverBankDecoderR10 C00

Pagina 64

Agilent E1465A Agilent E1466A Agilent E1467ACH # Relay RelayDriverBankDecoderCH # Relay RelayDriverBankDecoderCH # Relay RelayDriverBankDecoderR12 C00

Pagina 65

Agilent E1465A Agilent E1466A Agilent E1467ACH # Relay RelayDriverBankDecoderCH # Relay RelayDriverBankDecoderCH # Relay RelayDriverBankDecoderR14 C00

Pagina 66

Self-TestError CodesTable 4-4 shows the self-test error codes for the matrix modules. Themeaning of each code is given in the right-hand column. If a

Pagina 67

Declaration of Conformity Declarations of Conformity for this product and for other Agilent products may be downloaded from the Internet. There are

Pagina 68

Figure 4-2. E1465A/E1466A/E1467A Disassembly68 Service Agilent E1465A/66A/67A Service Manual

Pagina 69 - Error Codes

Repair/MaintenanceGuidelinesThis section provides guidelines for repairing and maintaining theAgilent E1465A/E1466A/E1467A matrix modules, including:•

Pagina 70

Post-Repair Safety ChecksAfter making repairs to the module, inspect the device for any signs ofabnormal internally generated heat, such as discolored

Pagina 71 - Maintenance

Notes6 Agilent E1465A, E1466A, E1467A Matrix Switch Modules Service Manual

Pagina 72 - Diagrams

NotesAgilent E1465A, E1466A, E1467A Matrix Switch Modules Service Manual 7

Comentarios a estos manuales

Sin comentarios