IntroductionIn many cases, power meters and power sensors are used for system calibration in manufactur-ing. During system calibration, frequency swee
2Frequency Sweep ModeFrequency Sweep mode is used in a frequency response calibration system where the ampli-tude is constant and the frequency of the
3Power Sweep ModePower Sweep mode is used in a power level calibration setup when the frequency is constant (CW frequency) and the amplitude of the po
4One-Way External Triggering CommunicationThe signal source’s TRIG OUT is connected to the power meter’s TRIG IN (see Figure 1). Then the Start, Stop,
5External triggering 2-way frequency sweep measurementUse the following SCPI commands for external triggering 2-way frequency sweep measurement.Table
6External triggering 2-way power sweep measurementUse the following SCPI commands for external triggering 2-way power sweep measurement.Table 4. SCPI
7Comparison of 1-way and 2-way External TriggeringIn 1-way or 2-way external trigger measurement, measurement data is stored in the buffer and the FE