Functions and Measurements 3
N9320A Spectrum Analyzer 57
9 Measure the rejection of the low pass filter:
Press
> Normal > 370 > MHz, Delta > 160> MHz
The marker readout displays the rejection of the filter at 160
MHz above the cutoff frequency of the low pass filter. See
Figure 19 as below.
Figure 19 Measure the Rejection Range