6 SCPI Command Reference
126 N9340A User’s Guide
Set reference level to the marker value
:CALCulate:MARKer[1]|2|3|4|5|6[:SET]:RLEVel
This command sets the reference level to the
specified marker amplitude. In delta marker mode,
the reference level is set to the amplitude
difference between the markers.
Marker On/Off
:CALCulate:MARKer[1]|2|3|4|5|6:STATe
OFF|ON|0|1
:CALCulate:MARKer[1]|2|3|4|5|6:STATe?
This command toggles the selected marker status
between on and off.
Marker to Trace
:CALCulate:MARKer[1]|2|3|4|5|6:TRACe <integer>
:CALCulate:MARKer[1]|2|3|4|5|6:TRACe?
This command assigns the specified marker to the
designated trace 1, 2, 3 or 4.
Key access: [Marker] > {Marker To} > {To Ref}
Key access: [Marker] > {Normal}
[Marker] > {Off}
*RST value: 1
Range: 1, 2, 3, or 4
Key access: [Marker] > {More} > {Marker Trace}