SCPI Command Reference 6
N9340A User’s Guide 151
Tracking Generator Subsystem
The tracking generator subsystem controls the use
of the tracking generator function of the analyzer.
State
:TGENerator:AMPLitude:STATe OFF|ON|0|1
:TGENerator:AMPLitude:STATe?
This command toggles the state of the tracking
generator between on and off. Setting it to the ON
state enables the tracking generator mode.
Output Amplitude
:TGENerator:AMPLitude <ampt>
:TGENerator:AMPLitude?
This command sets the tracking generator output
amplitude. This command is valid when the track-
ing generator is enabled.
*RST State: OFF
Key Access: 1 Press [MODE] to enable the tracking gen-
erator mode.
2 Press {Amplitude} to toggle the tracking
generator between on and off.
*RST State: –20 dBm
Range: 0 to –30 dBm
Available unit: dBm
Key Access: {Amplitude}