6 SCPI Command Reference
126 N9340A User’s Guide
[:SENSe]:BANDwidth Subsection
Resolution Bandwidth
[:SENSe]:BANDwidth|BWIDth [:RESolution]
<freq>
[:SENSe]:BANDwidth|BWIDth [:RESolution]?
This command specifies the resolution bandwidth.
Resolution Bandwidth Automatic
[:SENSe]:BANDwidth|BWIDth
[:RESolution]:Auto OFF|ON|0|1
[:SENSe]:BANDwidth|BWIDth
[:RESolution]:Auto?
This command couples the resolution bandwidth to
the frequency span. Auto- couple resolution
bandwidth is not available in zero span..
Video Bandwidth
[:SENSe]:BANDwidth|BWIDth:VIDeo <freq>
[:SENSe]:BANDwidth|BWIDth:VIDeo?
This command specifies the video bandwidth.
*RST Value: 1 MHz
Range: 30 Hz to 1 MHz
Available Unit: Hz, kHz, MHz, GHz
Key Access: [BW/SWP] > {RBW}
*RST State: On
Key Access: [BW/SWP] > {RBW}
*RST Value: 1 MHz
Range: 3 Hz to 1 MHz
Available unit: Hz, kHz, MHz, GHz
Key Access: [BW/SWP] > {VBW}