100 Agilent 34980A Mainframe User’s Guide
3 Features and Functions
Thermistor Measurements
The instrument supports 2.2 k (YSI Series 44004), 5 k (YSI Series 44007),
and 10 k
(YSI Series 44006) thermistors.
Front Panel Operation: To select the thermistor function for the active
channel, choose the following items.
DMM or Channel (Configure) > TEMPERATURE > PROBE TYPE > THERMISTOR
To select the thermistor type for the active channel, choose from the
following items.
THERMISTOR TYPE > 10K|5K|2.2K
Remote Interface Operation: You can use the CONFigure or MEASure?
command to select the probe type and thermistor type. For example,
the following command configures channel 3001 for measurements of a
5 k
thermistor:
CONF:TEMP THER,5000,(@3001)
You can also use the SENSe command to select the probe type and
thermistor type. For example, the following command configures channel
1003 for measurements of a 10 k
thermistor:
SENS:TEMP:TRAN:THERM:TYPE 10000,(@1003)