Repairing the Pump
Reassembling the Pump Head Assembly
Reassembling the Pump Head Assembly
Tools required:
4-mm hexagonal key
Pump head grease (79846-65501)
1 Place the support rings on the plunger
housing (plungers not installed) and snap the
pump head and plunger housing together.
2 Insert the plungers and carefully press them
into the seal.
3 Tighten the lock screw.
Pump head
Support ring
Plunger housing
Plunger
Lock screw