Agilent Technologies 8753ET Guía de usuario Pagina 254

  • Descarga
  • Añadir a mis manuales
  • Imprimir
  • Pagina
    / 546
  • Tabla de contenidos
  • MARCADORES
  • Valorado. / 5. Basado en revisión del cliente
Vista de pagina 253
1-240 Chapter 1
Alphabetical Command Reference
SPEC
SPEC
Syntax
SPEC<$><num,num,…,num>;
Description
The following commands initiate the SPECIFY CLASS part of modifying a cal kit. With
each command, send the analyzer a series of standard numbers to be included in the class.
When the class is full, send the CLAD command to terminate the specification.
Command Description Range for
<num,num,...>
Query Response
SPECFWDM Forward match standard numbers
1–8
N/A
SPECFWDT Forward transmission standard numbers
1–8
N/A
SPECRESP Response standard numbers
1–8
N/A
SPECRESI For Resp & Isol, specifies the response
standards
standard numbers
1–8
N/A
SPECREVM Reverse match standard numbers
1–8
N/A
SPECREVT Reverse transmission standard numbers
1–8
N/A
SPECS11A S11A standard numbers
1–8
N/A
SPECS11B S11B standard numbers
1–8
N/A
SPECS11C S11C standard numbers
1–8
N/A
SPECS22A S22A standard numbers
1–8
N/A
SPECS22B S22B standard numbers
1–8
N/A
SPECS22C S22C standard numbers
1–8
N/A
SPECTRLL TRL Line or Match standard numbers
1–8
N/A
SPECTRLT TRL Thru standard numbers
1–8
N/A
Vista de pagina 253
1 2 ... 249 250 251 252 253 254 255 256 257 258 259 ... 545 546

Comentarios a estos manuales

Sin comentarios