Agilent Technologies 8922S Especificaciones Pagina 163

  • Descarga
  • Añadir a mis manuales
  • Imprimir
  • Pagina
    / 244
  • Tabla de contenidos
  • MARCADORES
  • Valorado. / 5. Basado en revisión del cliente
Vista de pagina 162
12-33
Module I/O Specifications
A16 Receiver
Outputs
Figure 4-8 Typical Display
To test Pulse Demod apply RF Carrier with AM modulation to RF Input, measure Pulse
Demod Out on oscilloscope ≅ 180 mV x %MOD
NOTE: Ensure correct settings on RF Analyzer page (frequency/amplitude). If the RF Input level
is greater than 5 dBm below RF Analyzer setting, the measurement will not register.
To A9 Global Test and Demod
UNMUTED_FM J1(14)
Sensitivity into 100 k load:
20 µV/Hz See Figure 4-8 on
page 12-33
Max Deviation: 100 kHz peak
Amplitude 1.8 Vdc
To Front Panel
PULSE_DEMOD J1(7)
Level Pulse ON into open circuit: +2 V
Level Pulse OFF: 0 V
Output Impedance: 600
10-90% Rise/Fall time: <2.5 µs
Vista de pagina 162
1 2 ... 158 159 160 161 162 163 164 165 166 167 168 ... 243 244

Comentarios a estos manuales

Sin comentarios