Description of Screens 5
N9360A Multi UE Tester CDMA2000 User Manual 5-63
More (1 of 3): Displays the Softkey Menu 2
Softkey Menu 2
Print Screen: Prints a hardcopy of the screen or saves a
graphic file of the screen in a USB memory device
depending on the setting of the Printer input field on the
[Configuration] screen.
Softer Handoff: Starts Softer handoff.
Code Power: Displays the [Code Domain Power]
measurement screen.
More (2 of 3): Displays the Softkey Menu 3.
Softkey Menu 3
Debug Data > USB: This softkey is unavailable.
Call Drop Time On/Off: Sets Call Drop Timer to
On or Off.
Meas RX Power: Starts the RX Power measurement. This
softkey is unavailable when the measurement item of the
RF Power is set to
Off (not measured).
More (3 of 3): Returns to the Softkey Menu 1.
Screen Field
Set the input items required to execute tests, referring to the
description shown in Table 5-13.
Table 5-13 Input Items on [Manual Test] Connection Screen (MC-1x)
Input Field Description
Measuring
mode
Sets On (measured) or Off (not measured) for each test item group or measurement item.
TX Power
Frequency Error
Rho, Origin Offset or Multi code Rho, Origin Offset
Time Offset
FER or TDSO FER
When the Radio Config is set to F1R1 or F2R2, Rho and Origin Offset are displayed. When it is set to
F3R3, F4R3 or F5R4, Multi-code Rho and Origin Offset are displayed.
When the Radio Config is set to F3R3, F4R3 or F5R4, and the Service Option is set to 32, TDSO FER is
displayed. The other settings, FER is displayed.