Performance and Specifications 7
N9360A Multi UE Tester CDMA2000 User Manual 7-7
Waveform Quality (Rho/Multi-Code Rho) measurement
The specifications of Waveform Quality (Rho/Multi-Code Rho)
measurement are shown on Table 7-12.
Time Offset measurement
Time Offset Measurement specifications are shown on
Table 7-4.
Table 7-12 Waveform Quality (Rho/Multi-Code Rho) measurement
Item Specification Unit Remarks
Input Level –20 to +36 dBm
Measurement Range 0.900 to 0.999
Resolution 0.001
Residential Rho 0.990 Downlink Output Level
: <–50 dBm
Table 7-13 Time Offset Measurement
Item Specification Unit Remarks
Input Level –20 to +36 dBm
Measurement Range ±10 μsec
Resolution 1/8 chip (Approximately 102 nsec)