Page 27
安捷倫科技高頻元件量測研討會
Feb.23, 2006
Two Ways to Perform the Delta-Match Calibration
Global Delta Match
Covers entire frequency range of instrument
Interpolation used for TRL cals with smaller spans and/or less points
Only one global delta match cal set can exist at a time
Future: considering providing factory global delta match cal set
User Delta Match
Perform a non-TRL calibration, such as mechanical SOLT, SOLR or ECal
Save data as a normal user cal set
Frequency range and number of points must be identical to that of the follow-on
TRL calibration
Interpolation is not used during the calibration, but can be used for measurements
Page 28
安捷倫科技高頻元件量測研討會
Feb.23, 2006
Performing the Global Delta Match Cal
Default Settings:
Points: 1601
IF BW: 100 Hz
Freq: 300 kHz -
20.1 GHz
User cannot change
these settings!