Agilent Technologies Option H48 Multiport Test Set Z5623A Especificaciones Pagina 160

  • Descarga
  • Añadir a mis manuales
  • Imprimir
  • Pagina
    / 191
  • Tabla de contenidos
  • MARCADORES
  • Valorado. / 5. Basado en revisión del cliente
Vista de pagina 159
22
Page 43
安捷倫科技高頻元件量測研討會
Feb.23, 2006
Compromises of Traditional Methods
Swap equal adapters
Need phase matched adapters of different sexes (e.g., f-f, m-f)
Error results from loss and mismatch differences of adapters
Use characterized thru
Multi-step process
Need a non-insertable cal to measure S-parameters of characterized thru
Perform adapter removal cal
Accurate but many steps in calibration (two 2-port calibrations)
Add adapters after cal, then, during measurement…
Use port extensions – doesn’t remove adapter mismatch effects
De-embed adapters (S-parameters known) – similar to characterized thru
Page 44
安捷倫科技高頻元件量測研討會
Feb.23, 2006
Non-Insertable ECal Modules
ECal resolves many, but not all non-insertable situations
Vista de pagina 159