Chapter 8: Trigger the Oscilloscope with the Timing Analyzer
Set Up the Timing Analyzer to Trigger on the Glitch
Set Up the Timing Analyzer to Trigger on the
Glitch
The timing analyzer is used to trigger on the glitch because it has the
ability to capture a glitch with minimal setup.
1 Click on the Trigger tab.
2 Under the Trigger Functions tab, click on ‘Find glitch’.
3 Click Replace to replace the default trigger sequence with the
‘Find glitch’ trigger function.
4 Under the Trigger Sequence 1, click on the field to the far-right
of Find TCOUNT edge.