3-67
Eye-Diagram Analyzer Reference
Agilent 70820A Calibration
To View Hardware
Response Traces
1 Display the traces menu by pressing:
MENU, Traces
2 Turn on both trace 1 and trace 2. Set the input to trace 1 as channel 1. Set the
input to trace 2 to channel 2.
3 Display the 100 kHz filter response of stage 0 by pressing:
page 1 of 2, Calib, IF calib, cal debug, show, SHOW CAL
4 Scale the traces by pressing:
page 2 of 2, Scale
5 Select trace 1, and press:
AUTOSCALE
6 Select trace 2, and press:
AUTOSCALE