Agilent Technologies 71501D Manual de usuario Pagina 149

  • Descarga
  • Añadir a mis manuales
  • Imprimir
  • Pagina
    / 176
  • Tabla de contenidos
  • MARCADORES
  • Valorado. / 5. Basado en revisión del cliente
Vista de pagina 148
3-69
Eye-Diagram Analyzer Reference
Agilent 70820A Calibration
5 To view the data for channel 1, press:
CH1 CAL INFO
6 To view the data for channel 2, press:
CH2 CAL INFO
7 To view additional pages of information, continue pressing:
CH1 CAL INFO or CH2 CAL INFO
Performing a Self-Test
You can initiate an automatic self-test routine that checks the internal condi-
tion of all major circuits. The self-test takes approximately 30 seconds to run.
(This routine automatically runs whenever the microwave transition analyzer
is turned on.)
Perform the self-test by pressing:
MENU, page 1 of 2, Calib, SELF-TEST
Controlling the Calibrator Signal
For verification purposes, the Calibrator Output signal can be manually con-
trolled. This includes changing its amplitude and frequency. The signal has the
following characteristics:
The signals amplitude can be set to one of five positions: 0 through 4. Zero
represents the smallest available amplitude. Four represents the largest
amplitude. These settings do not correlate to any specific amplitude and may
vary between instruments.
Shape square wave
Frequency 153 Hz to 5 MHz
Amplitude 5 positions
Vista de pagina 148
1 2 ... 144 145 146 147 148 149 150 151 152 153 154 ... 175 176

Comentarios a estos manuales

Sin comentarios