Agilent Technologies E4438C Guía de usuario Pagina 48

  • Descarga
  • Añadir a mis manuales
  • Imprimir
  • Pagina
    / 54
  • Tabla de contenidos
  • SOLUCIÓN DE PROBLEMAS
  • MARCADORES
  • Valorado. / 5. Basado en revisión del cliente
Vista de pagina 47
40 Users Guide
Signal Studio for WiMAX E4438C Option H13
Troubleshooting
CAUTION
Do not use “Save As” when saving the hosts file; this file does not have a file
extension.
Software Problems
After downloading the waveform, the ESG displays error: Baseband Generator DAC over
range.
Cause: The number of frames and some baseband filter selections (such as NONE) can
result in a high crest factor, causing digital-to-analog converter (DAC) overflow.
Solution: Reduce the 'runtime scaling' value on the “Project View” on page 22.
After downloading the waveform, the ESG's front panel keys do not work.
Cause: Downloading the waveform places the ESG in remote operation mode, which
deactivates front panel operation.
Solution: On the ESG, press the Local key to restore front panel operation.
Hardware Problems
ESG Signal Generator:
1. Check the ESG for any error messages; if possible, correct any conditions indicated by
the messages.
2. Refer to the ESG's documentation for further troubleshooting.
3. Contact an Agilent Technologies Service Center. See Page 45 for contact information.
Computer:
1. Verify that the equipment meets the minimum requirements, as described in “System
Requirements” on page 3.
2. Refer to the computer's documentation for troubleshooting and repair.
Vista de pagina 47
1 2 ... 43 44 45 46 47 48 49 50 51 52 53 54

Comentarios a estos manuales

Sin comentarios