Troubleshooting
Troubleshooting Performance Test Problems
1-98
Pulse Modulation On/Off Ratio •Perform:
No adjustments
If failing, troubleshoot the following assemblies:
A18 Reference (fails at all frequencies)
A13 Output (fails only 4 GHz)
A14 Extended Frequency Output (Option 506)
(fails only > 4 GHz)
Burst Modulation On/Off Ratio Perform:
All Burst adjustments
If failing, troubleshoot the following assemblies:
A18 Reference (fails at all frequencies)
A13 Output (fails only 4 GHz)
A14 Extended Frequency Output (Option 506)
(fails only > 4 GHz)
Table 1-40 Troubleshooting Performance Test Problems
Performance test that is failing... Action to perform...