Troubleshooting
Troubleshooting Assembly Level Problems
1-53
1100 Power Supply
Run the Self-Test to verify that the A12 CPU is receiving the following power supplies: +10 V Ref,
6VRef, 5.2 V Ref, and the +9 V Ref.
If the power supply voltages are correct on the J15 and J5 connector pins indicated:
a. Ohm the connection between the connectors J8 and J15, or J7 and J5.
If the connection is good, replace the A12 CPU.
(Refer to A12 CPU” on page 2-30.)
If the connection is bad, replace the A23 Motherboard.
(Refer to A23 Motherboard” on page 2-52.)
If the power supply voltages are incorrect on the J15 and J5 connector pins indicated:
a. Check the origin of the power supply.
If the power supply voltages are good at the origin, replace the A23 Motherboard.
(Refer to A23 Motherboard” on page 2-52.)
If the power supply voltages are bad, troubleshoot the problem using the assembly’s
troubleshooting procedure.