Performance Tests
Overview of Performance Tests
3-5
Figure 3-2 The DUT Selection Window
7. In the Select Test Equipment and Tests window (Figure 3-3), create a list of test equipment that will verify
the signal generator’s performance:
a. In the Device Type list (item 6), select a device type that you want to add to the equipment list.
b. In the Model list (item 7), select the model of the device type that you want to add to the equipment list.
c. In the Available Test Equipment box (item 8), select the serial number of an instrument. There may be
more than one instrument available for any model. Make sure that the instrument calibration due date
has not passed. If the due date has passed, use another available instrument. Also, alert your system
administrator that the instrument is due for calibration.
d. Click Add (item 9) to add the instrument to the test equipment list.
To delete a single entry in the equipment list, click on the unwanted selection in the Selected Test
Equipment box and then click Remove (item 11).
To delete the entire list of equipment, click Clear (item 12).
e. Continue adding test equipment to the list by repeating these steps.