Error Messages
Agilent Technologies
E8257D/67D PSG Signal Generators
This guide applies to the following signal generator models:
E8257D PSG Analog Signal Generator
E8267D PSG Vector Signal Generator
Due to our continuing efforts to improve our products through firmware and hardware revisions, signal
generator design and operation may vary from descriptions in this guide. We recommend that you use the
latest revision of this guide to ensure you have up-to-date product information. Compare the print date of this
guide (see bottom of page) with the latest revision, which can be downloaded from the following website:
www.agilent.com/find/psg
Manufacturing Part Number: E8251-90357 - For Online Use Only
Printed in USA
August 2008
© Copyright 2001-2008 Agilent Technologies, Inc.
Vista de pagina 0
1 2 3 4 5 6 ... 70 71

Indice de contenidos

Pagina 1 - Error Messages

Error MessagesAgilent TechnologiesE8257D/67D PSG Signal GeneratorsThis guide applies to the following signal generator models:E8257D PSG Analog Signal

Pagina 2

lpErrorDoc.txt fragmentation). Report this error to Agilent Technologies. The instrument is still functional.-321 Out of memory;

Pagina 3 - Introduction

lpErrorDoc.txt Cannot restore specified cal from backup because of a file error.-313 Calibration memory lost; Cannot store cal to backup: {

Pagina 4 - Error Message Format

lpErrorDoc.txt valid state.-310 System error; Critical digital module error. Report this error to Agilent Technologies. This

Pagina 5 - Error Message Types

lpErrorDoc.txt The command issued expected a given personality to be loaded on the given board. Report this error to Agilent Te

Pagina 6

lpErrorDoc.txt File system volumes were verified. Informational warning. Powering down system (turning the power off) durin

Pagina 7

lpErrorDoc.txt The calibration will take an initial coarse YO pretune calibration that consists of a gain and offset term, and r

Pagina 8

lpErrorDoc.txt-271 Macro syntax error A syntactically legal macro program data sequence, written in accordance with IEEE 488.2,

Pagina 9

lpErrorDoc.txt-259 LAN setup error; IP address and default gateway must be on the same subnet. Offending entries are changed. The IP ad

Pagina 10

lpErrorDoc.txt-256 File name not found; {name} The burst shape type has been defaulted. The specified file does not exist, so the burst sha

Pagina 11

lpErrorDoc.txt properly. Report this error to Agilent Technologies.-230 Data corrupt or stale; Either the file is not a bit file, or the

Pagina 12

ii NoticeThe material contained in this document is provided “as is”, and is subject to being changed, without notice, in future editions.Further, to

Pagina 13

lpErrorDoc.txt-224 Illegal parameter value; The OSR must be an even value. RADio:ARB:GENerate:SINE generates a valid sine waveform and the

Pagina 14

lpErrorDoc.txt Due to hardware limitations, only a certain number of unique data points can be created for the IQ map. IQ rotati

Pagina 15

lpErrorDoc.txt command.-222 Data out of range; A burst rise shape must have at least two points. Two points are needed to specif

Pagina 16

lpErrorDoc.txt The sweep/list manual point has been reassigned to a smaller number value due to the longest list decreasing in

Pagina 17

lpErrorDoc.txt source frequency; number of points and start/stop values for list mode; start/stop values for internal I/Q calibr

Pagina 18

lpErrorDoc.txt turn the mode on. HSPA has been turned off. To output a valid RF signal, at least one channel must be turned

Pagina 19

lpErrorDoc.txt / NumberOfBlks. But block size does not match the BlockSetSize exactly. Try another value. Block size has been adjus

Pagina 20

lpErrorDoc.txt-221 Settings conflict; Cannot turn AWGN on, option 403 is required for any AWGN functionality Option 403 (AWGN) is requi

Pagina 21

lpErrorDoc.txt The clipping function is not available for the CDMA2000 or 1xEV-DV reverse link radio configurations 1 and 2. The

Pagina 22

lpErrorDoc.txt format is off: {name}. This error is reported when a file header command is tried when the format is not on.

Pagina 23

Error MessagesIntroductionIntroductionIf an error condition occurs in the signal generator, it is reported to both the front panel display error queu

Pagina 24

lpErrorDoc.txt This error is generated if the module is not connected when digital data is turned on.-221 Settings conflict; Th

Pagina 25

lpErrorDoc.txt E-TFCI value should be 0 to 125 when E-TFCI table is 1 and TTI is 2msec. E-TFCI value shoudl be 0 to 1

Pagina 26

lpErrorDoc.txt match the FM1/PM1 deviation.-221 Settings conflict; FM2/PM2 value set greater than FM1/PM1 value. FM1/PM1 changed to mat

Pagina 27

lpErrorDoc.txt-221 Settings conflict; The frequency spacing has been limited by the number of carriers. The frequency spacing and the n

Pagina 28

lpErrorDoc.txt-221 Settings conflict; In fixed position the block size must be either 0 or equal to block set size. The block size has been set to

Pagina 29

lpErrorDoc.txt-221 Settings conflict; Invalid Channel code. Channel code has been changed to the highest Channel code for this symbol rate

Pagina 30

lpErrorDoc.txt-221 Settings conflict; IR buffer size is too small. The 3GPP standard defines the transport block set size as 0 to

Pagina 31

lpErrorDoc.txt GAP2 is a single frame method gap that is greater than 7 slots long. DTX slots per frame should be less than 8.-2

Pagina 32

lpErrorDoc.txt data source is external. Single is not a valid repeat setting for an external data source.-221 Settings conflict; PHS or

Pagina 33

lpErrorDoc.txt-221 Settings conflict; QPCH and FFCH/OCNS cannot be on at the same time. FFCH/OCNS has been turned off. QPCH is generate

Pagina 34

Error MessagesError Message FormatError Message FormatWhen accessing error messages through the front panel display error queue, the error numbers, me

Pagina 35

lpErrorDoc.txt An attempt was made to setup the R-DCCH while in a radio configuration that does not support this channel. This c

Pagina 36

lpErrorDoc.txt channel, you must change to the proper radio configuration before continuing.-221 Settings conflict; RSCH is only

Pagina 37

lpErrorDoc.txt firmware release is not available, then an alternate solution is to load a compatible FPGA image by pressing the Update SSIO In

Pagina 38

lpErrorDoc.txt 15 >= TGSN + TGL2 + TGD TGD parameter is set to Undefined or 0. TGPL1 * 15 >= TGSN + TGL1 TGPL1 is the duratio

Pagina 39

lpErrorDoc.txt increased. To avoid the error, increase the symbol rate of physical channel or decrease the number of bits of Tr

Pagina 40

lpErrorDoc.txt is defaulted back to EDGE modulation.-221 Settings conflict; When Skew Mode is enabled, the maximum number of filter sym

Pagina 41

lpErrorDoc.txt The data rate is not contained in the defined data rate table.-220 Parameter error; Spread factor exceeds maximum allowable

Pagina 42

lpErrorDoc.txt An arming signal was received and recognized by the device but was ignored.-211 Trigger ignored A GET,

Pagina 43

lpErrorDoc.txt or type of parameters. -183 Invalid inside macro definition The program message unit sequence, sent with a *DDT

Pagina 44

lpErrorDoc.txt The specified filename exceeds the maximum length.-151 Invalid string data; Header Description String is too long and was

Pagina 45

Error MessagesError Message TypesError Message TypesEvents do not generate more than one type of error. For example, an event that generates a query e

Pagina 46

lpErrorDoc.txt 7.7.2.4.1).-123 Exponent too large The magnitude of an exponent was greater than 32000 (see IEEE 488.

Pagina 47

lpErrorDoc.txt-104 Data type error The parser recognized a data element that is not allowed. For example, numeric or string dat

Pagina 48

lpErrorDoc.txt In 8340/8757 compatibility mode, the OA query was received but it is not supported for the current active functio

Pagina 49

lpErrorDoc.txt211 Not owner; {name} Permission has been denied to perform an operation on the specified file.219 Bad file numbe

Pagina 50

lpErrorDoc.txt mode, the user has requested a power setting that is below the ALC range for the attenuator setting. The power ha

Pagina 51

lpErrorDoc.txt517 Calibration failure; Cal aborted by user. While executing the internal I/Q calibration, either a DCAS was sent

Pagina 52

lpErrorDoc.txt While executing the internal I/Q calibration, a failure occurred while attempting to access the IQ CAL CONSTS Arr

Pagina 53

lpErrorDoc.txt While executing the internal I/Q calibration, a failure occurred. An internal software error has occurred. A meas

Pagina 54

lpErrorDoc.txt While executing the internal I/Q calibration, a failure occurred. Please check the hardware.520 Sampler Loop unlo

Pagina 55

lpErrorDoc.txt One or more OCNS channels must be enabled for channel balancing to occur.600 RPP has tripped. The reve

Pagina 56

-440 Query UNTERMINATED after indefinite response A query was received in the same program message in which a query requesting a

Pagina 57

lpErrorDoc.txt to respond within the appropriate amount of time. If cycling power does not correct the problem, report this erro

Pagina 58

lpErrorDoc.txt During the system configuration scan of parallel bus boards, a board with a bad configuration header was detected

Pagina 59

lpErrorDoc.txt file conflict or a hardware configuration error. Report this error to Agilent Technologies.617 Configuration Erro

Pagina 60

lpErrorDoc.txt errors below 3.2 GHz. The lowband coupler detector has failed. Report this error to Agilent Technologies.625

Pagina 61

lpErrorDoc.txt provide a challenge value; this value cannot be zero.627 Licensing error; The waveform file already has or does not require

Pagina 62

lpErrorDoc.txt First {name} is file name, second is file revision and third is supported revision.629 File format invalid; Seque

Pagina 63

lpErrorDoc.txt begin on an even sample pair boundary.630 Arb Sequence Error; Unable to reset sequence on baseband board after change in

Pagina 64

lpErrorDoc.txt time. This error is reported when the user attempts to activate one type of modulation while the other is activa

Pagina 65

lpErrorDoc.txt messages may clarify why this error was reported.700 State Save Recall Error; Failed to save the instrument state. {name}

Pagina 66

lpErrorDoc.txt Burst shape files must have two or more points for both the rise and the fall shape specification.702 Load/Store

Pagina 67

lpErrorDoc.txt-321 Out of memory; ARB is not functional. There is insufficient memory for the ARB to generate the requested wave

Pagina 68

lpErrorDoc.txt Digital module: {name}. Report this error to Agilent Technologies. This error is reported when the FPGAs on the digital

Pagina 69

lpErrorDoc.txt valid, but old version of the interface FPGA.808 Digital module unlock error; The output clock is not locked. The clock

Pagina 70

lpErrorDoc.txt A file system operation has failed due to a lack of memory in the system. Reduce memory usage and try again.-321

Pagina 71

lpErrorDoc.txt There is not enough memory to read the specified data file. Try a smaller file. Reduce the size of any list/sweep