150 Agilent N516xA, N518xA, E8663B, E44x8C, and E82x7D Signal Generators Programming Guide
Programming the Status Register System
Overview
registers using summary bits. The Status Byte Register is at the top of the hierarchy and contains the
status information for lower level registers. The lower level registers monitor specific events or
conditions.
The lower level status registers are grouped according to their functionality. For example, the Data
Questionable Frequency Status Group consists of five registers. This chapter may refer to a group as
a register so that the cumbersome longer description is avoided. For example, the Standard Operation
Status Group’s Condition Register can be referred to as the Standard Operation Status register. Refer
to “Status Groups” on page 169 for more information.
Figure 4- 1, Figure 4- 2, Figure 4- 3, and Figure 4- 4, Figure 4- 5, Figure 4- 6, Figure 4- 7, and Figure 4-8
shows each signal generator model’s the signal generator’s status byte register system and hierarchy.
The status register systems use IEEE 488.2 commands (those beginning with *) to access the
higher- level summary registers (refer to the SCPI Reference). Access Lower- level registers by using
STATus commands.