18
Specifications
Two-tone
IM distortion
250 kHz to 3.2 GHz <–45 dBc for RF levels < 0dBm (typical)
>3.2 GHz to 20 GHz <–55 dBc for RF levels < 0dBm (typical)
Measured two-tone
Multitone
Signal Studio for Enhanced Multitone
There is optional software for the E8267C PSG vector signal generator to improve the signal
quality of two-tone and multitone test signals. Signal Studio for enhanced multitone,
Option 408 for the E8267C, significantly improves the IMD suppression, RF flatness,
image suppression, and carrier feed through performance of multitone test signals
generated by the PSG. The improved signal quality is achieved by applying pre-distortion
to the I/Q waveform based on an iterative correction algorithm. Refer to the Signal Studio
for Enhanced Multitone Technical Overview (publication number 5988-5639EN) for
additional information on this application.
Ref-4 dBm
Samp
Log
10
dB/
LgAv
10
W1 S2
S3 FC
(f):
#f>50K
#Swp
#Atten 6 dB
Mkr1 -20.00 MHz
-62.275 dB
Center 20.000 00 GHz
Res BW 470 kHz
VBW 470 kHz
Span 50 MHz
Sweep 1 ms
Marker
-20.000000 MHz
-62.275 dB
Ref 0 dBm
Norm
Log
10
dB/
LgAv
W1 S2
S3 FC
(f):
#f>50K
#Swp
Atten 10 dB
Mkr1 -38.67 MHz
-41.19 dB
Center 20.000 00 GHz
#Res BW 3 kHz
VBW 3 kHz
Span 80 MHz
Sweep 10.72 s
Marker
-38.670000 MHz
-41.19 dB
Ref-4 dBm
Samp
Log
10
dB/
LgAv
10
W1 S2
S3 FC
(f):
#f>50K
#Swp
#Atten 6 dB
Mkr1 -8.00 MHz
-59.104 dB
Center 20.000 00 GHz
Res BW 180 kHz
VBW 180 kHz
Span 20 MHz
Sweep 2.36 ms
Marker
-8.000000 MHz
-59.104 dB
Measured multitone Measured image rejection > 3.2 GHz