Agilent Technologies EXG Manual de usuario Pagina 10

  • Descarga
  • Añadir a mis manuales
  • Imprimir
  • Pagina
    / 31
  • Tabla de contenidos
  • MARCADORES
  • Valorado. / 5. Basado en revisión del cliente
Vista de pagina 9
10
Maximum reverse power, nominal
< 1 GHz 50 W
> 1 to < 2 GHz 25 W
> 2 to < 6 GHz 20 W
Max DC voltage 50 VDC
Trip level 2 W
Amplitude switching speed
1
Standard Option UNZ Option UNZ, typical
CW mode
SCPI mode ≤ 5 ms, typical ≤ 750 µs ≤ 650 µs
Power search SCPI mode < 12 ms, measured
List/step sweep mode ≤ 5 ms, typical ≤ 500 µs ≤ 300 µs
Digital modulation on (N5172B only)
SCPI mode ≤ 5 ms, typical ≤ 1.15 ms ≤ 950 µs
Power search SCPI mode < 12 ms, measured
List/step sweep mode ≤ 5 ms, typical ≤ 900 µs ≤ 400 µs
Alternate power level control (N5172B only)
Switching time
(via waveform markers)
20 µs within ± 1 dB, measured
Functional power range –15 dBm to –144 dBm, measured
User flatness correction
Number of points 3201
Number of tables Dependent on available free memory in instrument; 10,000 maximum
Entry modes USB/LAN direct power meter control, LAN to GPIB and USB to GPIB, remote bus and manual
USB/GPIB power meter control
Sweep modes
See Frequency Specifi cations section for more detail
1. Time from receipt of SCPI command or trigger signal to amplitude settled within 0.2 dB. Switching speed specifi cations apply when status register