Agilent Technologies EXG Guía de usuario Pagina 119

  • Descarga
  • Añadir a mis manuales
  • Imprimir
  • Pagina
    / 426
  • Tabla de contenidos
  • SOLUCIÓN DE PROBLEMAS
  • MARCADORES
  • Valorado. / 5. Basado en revisión del cliente
Vista de pagina 118
Agilent X-Series Signal Generators User’s Guide 105
Optimizing Performance
Using Flatness Correction
Recalling and Applying a User Flatness Correction Array
The following example assumes that a user flatness correction array has been created and stored. If
not, perform the Example: A 500 MHz to 1 GHz Flatness Correction Array with 10 Correction
Values on page 102.
1. Preset the signal generator.
2. Recall the desired User Flatness Correction file:
a. Press AMPTD > More > User Flatness > Configure Cal Array > More > Preset List > Confirm Preset.
b. Press Load/Store.
c. Highlight the desired file.
d. Populate the user flatness correction array with the data contained in the selected file:
Press Load From Selected File > Confirm Load From File.
The user flatness correction array title displays User Flatness: Name of File.
3. Apply the correction data in the array to the RF output: Press Return > Flatness Off On to On.
Vista de pagina 118
1 2 ... 114 115 116 117 118 119 120 121 122 123 124 ... 425 426

Comentarios a estos manuales

Sin comentarios