Agilent N5161A/62A/81A/82A/83A MXG Signal Generators User’s Guide 217
Basic Digital Operation (Option 651/652/654)
Modifying a FIR Filter Using the FIR Table Editor
Figure 8-24
7. Press Return.
Modifying the Coefficients
1. Using the front panel arrow keys, highlight coefficient 15.
2. Press 0 > Enter.
3. Press Display Impulse Response.
Figure 8-25
Refer to Figure 8- 25 on page 217. The graphic display can provide a useful troubleshooting tool
(in this case, it indicates that a coefficient value is missing, resulting in an improper Gaussian
response).
4. Press Return.
5. Highlight coefficient 15.
6. Press 1 > Enter.
For details on each key, use key
help as described on page 42.
For details on each key, use key
help as described on page 42.