Agilent Technologies E1356A Especificaciones Pagina 118

  • Descarga
  • Añadir a mis manuales
  • Imprimir
  • Pagina
    / 149
  • Tabla de contenidos
  • MARCADORES
  • Valorado. / 5. Basado en revisión del cliente
Vista de pagina 117
Procedure Measure the guard voltage on internal channel 14.
MEASure:VOLT:DC? (@114)
The following example measures the guard voltage. The program assumes
the strain gage multiplexer is used with the HP E1326B/E1411B multimeter
as part of a scanning multimeter instrument. The computer syntax used is
HP BASIC running on a Hewlett-Packard Series 300 computer.
!Clear and reset the multimeter.
CLEAR 70903
OUTPUT 70903;"*RST"
!Measure the guard voltage on internal channel 14.
OUTPUT 70903;"MEAS:VOLT:DC? (@114)"
ENTER 70903;G_volt
!Print the guard voltage.
PRINT “Guard voltage = ;G_volt
END
A typical output from this program is:
Guard voltage = 2.5233
Appendix C Strain Gage Diagnostics 117
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com
Vista de pagina 117
1 2 ... 113 114 115 116 117 118 119 120 121 122 123 ... 148 149

Comentarios a estos manuales

Sin comentarios