Agilent Technologies C-Size VXIbus Systems Manual de usuario Pagina 3

  • Descarga
  • Añadir a mis manuales
  • Imprimir
  • Pagina
    / 8
  • Tabla de contenidos
  • MARCADORES
  • Valorado. / 5. Basado en revisión del cliente
Vista de pagina 2
3
EMC Filler Panel (1-slot wide)
The Agilent E8400-60202 filler panel (1-slot wide) provides
continuous EMC connection across the mainframe front
opening. This helps reduce radiated emissions and
increases ESD immunity. All Agilent VXI modules include
EMC contacts to adjacent slots.
VXI Slot Filler Panel (1-slot wide)
The E8400-44305 VXI Slot Filler Panel is 1-slot wide and is
used to cover unused slots, providing a smooth appearance
and preventing access. In addition to the cosmetic and
safety value, the slot filler panel suppresses acoustic noise
and improves airflow. The E8400-44305 does not include
EMC contacts.
Air Flow Restricter (1-slot)
The one-slot E8400-80917 Air Flow Restricter blocks airflow
in one empty slot.
Intermodule Chassis Shield Kit (1-slot wide)
The E8400-80919 Chassis Shield Kit (1-slot wide) provides
additional shielding between noisy modules. These shields
are relatively easy to install.
Documentation
Mainframe documentation consists of a User/Service
Manual which describes electrical and mechanical aspects
of the mainframe and accessories. The manual is included
on Agilent’s Universal Instrument Drivers CD-ROM supplied
with the mainframe and is also available on Agilent’s
Website. A printed copy of the manual is also included.
Warranty
Agilent Technologies provides a standard 3-year return-to-
Agilent warranty on this mainframe. E8408A Opt. W01
converts the standard warranty to 1-year On-Site.
Product Specifications
Mechanical
Mainframe height:
(horizontal orientation)
Without bumpers
133 mm (5.25 in.)
(3 EIA rack units)
Mainframe width: 362 mm (14.3 in.)
Mainframe depth: 540 mm (21.3 in.)
Mainframe weight: 8.6 kg (19 lbs.)
Mainframe height:
(horizontal orientation)
With bumpers
152 mm (6.0 in.)
(3 EIA rack units)
Mainframe width: 388 mm (15.3 in.)
Mainframe depth: 548 mm (21.6 in.)
Mainframe weight: 9.1 kg (20 lbs.)
Available Current
Voltage
Peak Current I
MP
1
(Amps)
From 0 - 55° C:
2
Dynamic Current I
MD
1
(Amps)
From 0- 55° C:
3
+5V:
4, 5
17 A 2.5 A
+12V:
6
2.7 A 1.0 A
–12V:
6
2 A 0.8 A
+24V: 2 A 1.0 A
–24V: 2 A 1.0 A
–5.2V:
5
2 A 1.0 A
–2V: 2 A 1.0 A
1. Specifications apply at the mainframe backplane, 0 - 55° C.
2. I
MP
= Rated mainframe peak dc output current as defined by the VXIbus
Specification.
3. I
MD
= Rated mainframe peak-to-peak dynamic current as defined in the VXIbus
Specification by a current vs. frequency curve.
4. +5V must have a 1.0 A minimum load for other supplies to be in regulation
specification.
5. Maximum +5V current can be increased if -5.2V load is less than 2.0 A maximum or
-2V load is less than 2.0 A maximum. Maximum = 17.0 A + 1.2*(2.0 A - Load on -5.2V) +
0.4*(2.0 A - Load on -2V).
6. Maximum +12V current can be increased if -12V load is less than 2.0 A maximum.
Maximum = 2.7 A + 1.2*(2.0 A - Load on -12V).
Power
Temperature range: 0 to 40° C 40 to 50° C 50 to 55° C
Usable power
(90-264 Vac): 175 W 150 W 125 W
Power Input
Input voltage/frequency: 90 Vac min. - 140 Vac max.:
47 Hz min. - 440 Hz max.
90 Vac min. - 264 Vac max.:
47 Hz min. - 66 Hz max.
280 VA max.
Inrush current:
At input voltage 132 Vac: <35 A typical
At input voltage 264 Vac: <75 A typical
Power Switch
On/Standby switch on front panel.
Power Supply Protection
All voltages are protected from over-temperature, over-
voltage, short-to-ground and short-to-other-output.
Acoustic Noise
(Sound power at bystander position one meter in front of mainframe)
High-speed fan: <40 dBA
Variable fan on low speed: <30 dBA
Vista de pagina 2