Agilent Technologies DisplayPort Manual de usuario Pagina 13

  • Descarga
  • Añadir a mis manuales
  • Imprimir
  • Pagina
    / 16
  • Tabla de contenidos
  • MARCADORES
  • Valorado. / 5. Basado en revisión del cliente
Vista de pagina 12
13
Tests performed
The U7232A DisplayPort electrical
performance validation and compliance
software performs all the required
tests and many of the informative ones
as listed in the DisplayPort Compliance
Test Specification 1.1 (CTS1.1) as
documented in section 3.5.3 of the
base specification.
Note that the U7232A DisplayPort
test software can make other
measurements not included here, such
as receiver eye measurements and
equalized eye measurements as well
as most of the newly omitted
CTS 1.1 measurements.
Assertion
number
Test
description
Compliance
status
Included with
U7232A
Section 3
Test ID 3.1 Data eye diagram Normative Yes
Test ID 3.2 Non pre-emphasis level
verification
Normative Yes
Test ID 3.3 Pre-emphasis level verification Normative Yes
Test ID 3.4 Inter-pair skew Normative Yes
Test ID 3.5 Intra-pair skew Normative Yes
Test ID 3.6 Differential transition time Omitted* Yes
Test ID 3.7 Single-ended rise and fall
mismatch
Omitted* Yes
Test ID 3.8 Overshoot and undershoot Omitted* No
Test ID 3.9 Frequency accuracy Omitted* No
Test ID 3.10 AC common mode noise Normative Yes
Test ID 3.11 Non-ISI jitter measurements Normative Yes
Test ID 3.12 Total jitter measurements Normative Yes
Test ID 3.13 Unit interval Omitted* Yes
Test ID 3.14 Main link frequency compliance Omitted* Yes
Test ID 3.15 Spread spectrum modulation
frequency
Normative Yes
Test ID 3.16 Spread spectrum modulation
deviation
Normative Yes
Test ID 3.17 Spread spectrum dF/dt Informative No
The U7232A DisplayPort electrical
performance validation and compliance
software is compatible with Agilent
X90000 Series oscilloscopes with
operating software revision 2.50 or
higher, 90000 Series oscilloscopes
with operating software revision 1.1
or higher or the Agilent 80000 Series
oscilloscopes with operating software
revision A.05.40 or higher (Windows
®
XP Pro). For oscilloscopes with earlier
software revisions, free upgrade
software is available at http://www.
agilent.com/find/infiniium_software.
Agilent oscilloscopes for DisplayPort compliance test*
Suitable oscilloscopes Recommended new purchase Oscilloscope bandwidth
DSO80804A/B
DSO81004A/B
DSO812004A/B
DSO813004A/B
DSO/DSAX91604A
DSO/DSAX92004A
DSO/DSAX92504A
DSO/DSAX92804A
DSO/DSAX93204A
DSO90804A or DSA90804A
††
DSO91204A or DSA91204A
DSO91304A or DSA91304A
8 GHz*
10 GHz
12 GHz
13 GHz
16 GHz
20 GHz
25 GHz
28 GHz
32 GHz
* 8 GHz Bandwidth is recommended by the DisplayPort Compliance Test Specification.
The U7232A software will automatically control the bandwidth setting of the oscilloscope if the N5403A noise reduction software is
enabled.
†† The DSO80000 and DSO90000 Series oscilloscopes are upgradable in bandwidth from 2 to 13 GHz.
DSA models come with jitter and serial data analysis packages.
††† The Infiniium 90000 X-Series oscilloscopes are upgradeable in bandwidth from 16 to 32 GHz
Oscilloscope compatibility
* These tests were defined for DisplayPort CTS 1.0 and were informative. They have been deleted
for DisplayPort CTS 1.1
Table 2. DisplayPort electrical tests performed by the U7232A software
Table 1. Recommended oscilloscopes and bandwidth for the DisplayPort 1.1 specification