9
In addition to providing you with
measurement results, the U7232A
DisplayPort electrical test software
provides a report format that shows
you not only where your product
passes or fails, but also reports how
close you are to the limits specified for
a particular test assertion. You select
the margin test report parameter,
which means you can specify the
level at which warnings are issued
to alert you to the electrical tests
where your product is operating
close to the official test limit defined
by the DisplayPort compliance test
specification for a given test assertion.
Eye margin: Another method of
evaluating the eye is provided by
sweeping the eye horizontally with
the specified mask and determining
the last locations on the left and right
portions of the interior eye where
there are no violations in the mask.
The distance from the center mask
location is a measure of design margin
and is reported as eye margin if that
eye mask mode is chosen in the
configuration screen (Eye Diagram
Mask Movement: Find Margin – see
Figure 10 for example.)
Figure 12. The DisplayPort electrical test software results report documents your test,
indicates the pass/fail status, the test specification range, the measured values, and the
margin.
Margin is calculated:
Single-sided specification:
Margin = (Value
Specification
– Value
actual
)/Value
Specification
Double-sided specification:
Margin = lowest of: (Value
Specification_High
– Value
actual
)/Range
Specification
and (Value
Specification_Low
– Value
actual
)/Range
Specification
Margin analysis