2-12
Performance Verification
Front Panel Cal Signal
5 Subtract the second reading from the first reading, then divide the result by 4.
For example, if the first reading is +1.980 V and the second reading is -1.970 V, then:
Record the result in the Performance Test Record.
= 0.9875
+1.980 V - (-1.970 V)
4 V