4-15
Replacing Instrument Assemblies
To Remove the Display Backlights
To Remove the Display Backlights
1 Follow the instructions for removing the front panel on page 4-7, and the display on page 4-14.
2 Disconnect the cables from the A3 Backlight Inverter shown in Figure 4-18.
Figure 4-18. Disconnecting the Cables from the A3 Backlight Inverter
3 At the top of the display, push the locking tab to release the top backlight, as shown in Figure 4-19.
Figure 4-19. Release the Backlight Tab
4 Remove the top backlight, as shown in Figure 4-20.