Introduction to Programming
Duplicate Mnemonics
Duplicate Mnemonics
Identical function mnemonics can be used in more than one subsystem. For
example, the function mnemonic RANGE may be used to change the vertical
range or to change the horizontal range:
:CHANNEL1:RANGE .4
sets the vertical range of channel 1 to 0.4 volts full scale.
:TIMEBASE:RANGE 1
sets the horizontal time base to 1 second full scale.
CHANNEL1 and TIMEBASE are subsystem selectors and determine which
range is being modified.