Agilent Technologies G1978B Guía de usuario Pagina 12

  • Descarga
  • Añadir a mis manuales
  • Imprimir
  • Pagina
    / 72
  • Tabla de contenidos
  • MARCADORES
  • Valorado. / 5. Basado en revisión del cliente
Vista de pagina 11
12 Multimode Source for 6100 Series Single Quad LC/MS User Guide
1 Basic Operation and Maintenance
To create a method for alternating ESI and APCI operation
9 Click the OK button.
Figure 3 MM-ES and MM-APCI switching using Set Up MSD Signals dialog box
NOTE
In general, use the mixed mode operation (MM-ES+APCI setting with Signal 1) instead of
alternating MM-ES (Signal 1) and MM-APCI (Signal 2) mode switching. Two times as many
scans are obtained during elution of a chromatographic peak, and no delay is needed
between scans, resulting in better data. It is rarely necessary to know whether a compound
responds purely in ESI or APCI modes on a chromatographic time scale.
Ionization mode for
Signal 1 and Signal 2
Vista de pagina 11
1 2 ... 7 8 9 10 11 12 13 14 15 16 17 ... 71 72

Comentarios a estos manuales

Sin comentarios