Agilent-technologies N2610A Manual de usuario Pagina 107

  • Descarga
  • Añadir a mis manuales
  • Imprimir
  • Pagina
    / 248
  • Tabla de contenidos
  • MARCADORES
  • Valorado. / 5. Basado en revisión del cliente
Vista de pagina 106
Fiber Testing 4
Calibrating the Tester
Agilent FrameScope 350 User’s Manual 99
7 When the message below the diagram reads, “Ready to Calibrate,
press OK. The Calibration window opens.
8 When the window reads, “Calibration Successful,” press OK in the
window. The Fiber Main Menu displays. Remove the reference
jumper to start testing.
Calibrating for Single-Ended Testing
1 Insert a Fiber SmartProbe
+
in the FrameScope,.
2 Connect a test jumper to the Fiber SmartProbe
+
.
3 Connect the two ends of the test jumper to each other, using a
coupler.
4 Turn on the FrameScope.
The FrameScope detects a Fiber SmartProbe
+
, and displays the
following window:
5 On the Fiber Main Menu, press Calibration. The Fiber Calibration
screen displays.
Vista de pagina 106
1 2 ... 102 103 104 105 106 107 108 109 110 111 112 ... 247 248

Comentarios a estos manuales

Sin comentarios