3 Cable Testing with the FrameScope 350
Running an Autotest
Running an Autotest
To run an autotest:
1 If this is the first test of the day or after a long break, calibrate the
tester. See “Calibrating the Tester” on page 61.
2 Connect the FrameScope 350 and Dual Remote 350 to the ends of the
circuit needing certification.
3 On the FrameScope Main Menu, press the button identifying the
selected test. The test executes.
4 If the autotest passes, the screen will look like this:
5 To save the test results, see Saving Test Resultson page 79