Agilent Technologies 4000 X Manual de servicio Pagina 112

  • Descarga
  • Añadir a mis manuales
  • Imprimir
  • Pagina
    / 168
  • Tabla de contenidos
  • MARCADORES
  • Valorado. / 5. Basado en revisión del cliente
Vista de pagina 111
112 4000 X-Series Oscilloscopes Advanced Training Guide
4 Serial Bus Decoding & Triggering, Search & Navigation, and Segmented Acquisition Labs
You should now see LIN protocol decoding of this signal on your scope’s
display similar to Figure 74, but we still haven’t established stable
triggering. The scope is still triggering on any rising edge of channel-1
(default trigger condition). Let’s now set up the scope to trigger on a LIN
frame with the ID = 21
HEX
.
12 Press the [Trigger] front panel key.
13 Select Serial 1 (LIN) using the Trigger Type menu.
14 Tap Trigger on: Sync; then select ID – Frame ID.
15 Tap the Frame ID softkey; then set ID = 0x21 using the Entry knob.
Figure 74 Decoding a LIN bus while triggering on any rising edge of the input signal.
Vista de pagina 111
1 2 ... 107 108 109 110 111 112 113 114 115 116 117 ... 167 168

Comentarios a estos manuales

Sin comentarios