Agilent Technologies 4000 X Manual de servicio Pagina 40

  • Descarga
  • Añadir a mis manuales
  • Imprimir
  • Pagina
    / 168
  • Tabla de contenidos
  • MARCADORES
  • Valorado. / 5. Basado en revisión del cliente
Vista de pagina 39
40 4000 X-Series Oscilloscopes Advanced Training Guide
2 Oscilloscope Familiarization Labs
In addition to stopping acquisitions when an error is detected as shown in
Figure 21, you can also save a waveform, save an image, print, or perform
specific measurements when an error is detected. Using the “Run Until”
selection, you can also set up mask testing to run for a specific number of
tests, minimum time, or minimum Sigma test criteria. Note that in
addition to creating a pass/fail mask using the auto- mask creation feature
as demonstrated in this lab, you can also import a mask based on industry
specifications, such as an eye- diagram mask, from a USB memory device.
These types of imported masks can be easily created within a standard
text editor, such as Notepad on your PC.
Figure 21 Establishing specific test criteria.
Vista de pagina 39
1 2 ... 35 36 37 38 39 40 41 42 43 44 45 ... 167 168

Comentarios a estos manuales

Sin comentarios