Agilent Technologies 4000 X Manual de servicio Pagina 63

  • Descarga
  • Añadir a mis manuales
  • Imprimir
  • Pagina
    / 168
  • Tabla de contenidos
  • MARCADORES
  • Valorado. / 5. Basado en revisión del cliente
Vista de pagina 62
Advanced Triggering, Search & Navigation, and Segmented Acquisition Labs 3
4000 X-Series Oscilloscopes Advanced Training Guide 63
Your scope’s display should look similar to Figure 38 showing a positive
runt pulse that is approximately 100 ns wide near center- screen. In
addition to triggering on runt pulse conditions, the Agilent 4000 X-Series
oscilloscopes can also perform Search & Navigation to find multiple runt
pulse conditions, regardless of the specific trigger setup condition. Let’s
now capture a longer stream of pulses and perform an automatic search
operation to find all runts.
15 Set the scope’s timebase to 100.0 µs/div.
16 Press [Run/Stop] to stop repetitive acquisitions.
17 Press the [Search] front panel key.
18 Tap the Search Edge softkey; double- tap on Runt.
19 Tap the Settings softkey to observe the default runt search conditions.
Figure 38 Triggering on a positive runt pulse that is < 200 ns wide.
Vista de pagina 62
1 2 ... 58 59 60 61 62 63 64 65 66 67 68 ... 167 168

Comentarios a estos manuales

Sin comentarios