c. Move cursor to the bit assignment field and turn on appropriate eight
bits to be tested (* = on; . = off) as in figure below.
5. Set up the State Trace Specification without sequencing levels and set
Count States as in next figure.
a. Press front-panel TRACE key.
b. Move cursor to Count, press SELECT, then move cursor to States,
press SELECT, and set to any state by pressing SELECT again.
Figure 3-7. Bit Assignment for Data Test 1
Figure 3-8. Trace Specification for Data Test 1