96 Chapter2
Making Measurements
Example 7: Stimulus-Response Measurements
Figure 2-42 NORM REF LVL is a trace function.
After returning
NORM REF LVL to 0 dB, increase RANGE LVL to 30 dB. As
shown in Figure 2-43, the trace moves fully within the graticule.
Compare the settings: (1) input attenuator value has changed to 40 dB,
(2) the marker-amplitude readout displays 6.3 dB,and (3) the ERR 903
A>DLMT error message no longer appears.
Figure 2-43
RANGE LVL adjusts analyzer for compression-free measurements.