Agilent 81133A / 81134A Performance Verification
Agilent 81133A / 81134A Performance Verification, December 2003
21
4 Switch on the channels and check the low level at the following
settings (Program the High Level at least 500 mV larger than the Low
Level to be tested)
Low Level Acceptable Range TR Entry
- 2.00 V - 2.06 V to -1.94 V 5-1
-1.00 V - 1.04 V to -0.96 V 5-2
0 V - 20 mV to + 20 mV 5-3
1.00 V 0.96 V to 1.04 V 5-4
2.00 V 1.94 V to 2.06 V 5-5
5 Set up a "HI" pattern as follows: