Agilent Technologies 81134A Manual de usuario Pagina 9

  • Descarga
  • Añadir a mis manuales
  • Imprimir
  • Pagina
    / 35
  • Tabla de contenidos
  • MARCADORES
  • Valorado. / 5. Basado en revisión del cliente
Vista de pagina 8
Agilent 81133A / 81134A Performance Verification
Agilent 81133A / 81134A Performance Verification, December 2003
9
Test 2: Width
Range: 100 ps to (period - 100 ps)
Resolution: 1 ps
Accuracy: ± 40 ps after autocal within a temperature range of ± 10°
Digitizing Oscilloscope with Accessories
Cables 50 , SMA
1 Connect the 81133A/81134A Trigger Output to the Oscilloscope
Trigger Input, using a 20 dB attenuator
2 Connect the 81133A/81134A Channel 1 Output to the Oscilloscope
Channel 4 Input, using a 20 dB attenuator
3 Set up the Pulse Generator as follows:
Equiopment Needed
Procedure
Vista de pagina 8
1 2 3 4 5 6 7 8 9 10 11 12 13 14 ... 34 35

Comentarios a estos manuales

Sin comentarios