Agilent 81133A / 81134A Performance Verification
Agilent 81133A / 81134A Performance Verification, December 2003
25
Test 6: Transition Times
Transition Time 10% - 90% < 90 ps
Transition Time 20% - 80 % < 60 ps
Digitizing Oscilloscope with Accessories
Cable SMA, 50
1 Connect the 81133A/81134A Trigger Output to the Oscilloscope
Trigger Input, using a 20 dB attenuator
2 Connect the 81133A/81134A Channel Output to the Oscilloscope
Channel 3 input, using 20 dB attenuator
3 Set up the 81133A/81134A as follows and switch on the channel
Equipment Needed
Procedure