Agilent Technologies 81134A Guía de usuario Pagina 55

  • Descarga
  • Añadir a mis manuales
  • Imprimir
  • Pagina
    / 92
  • Tabla de contenidos
  • SOLUCIÓN DE PROBLEMAS
  • MARCADORES
  • Valorado. / 5. Basado en revisión del cliente
Vista de pagina 54
Setting Up a Repetitive Burst Signal Setting Up Generic and Advanced Signals
Agilent 81133A/81134A Pulse Generator User’s Guide, September 2008 55
Reconnect the DUT Enable the outputs:
Generated Signal Use the pulse generator’s TRIGGER OUT to trigger a scope.
To set up the trigger output:
1 On the Aux page, in the Trigger Output panel:
Select the trigger mode Pulse.
One trigger pulse occurs at the start of the repetitive data
pattern.
Set the Ampl to 1 V.
Set the Offset to 0 V.
Set the Divider to 256.
2 Enable the trigger output.
The Aux page looks now as follows:
Vista de pagina 54
1 2 ... 50 51 52 53 54 55 56 57 58 59 60 ... 91 92

Comentarios a estos manuales

Sin comentarios